GM700SB Specifications: Test strip: GS700, Coding : Auto-coding, Enzyme / Measurement technology: GDH / Electrochemical sensor, Blood sample source: Capillary, venous, arterial and neonatal whole blood, Minimum sample volume: 0.75 µL, HCT range: Glucose concentrations ≦ 11.1 mmol/L (200 mg/dL), HCT range 20 - 70% Glucose concentrations > 11.1 mmol/L (200 mg/dL), HCT range 20 - 60%. Measuring range: 0.6 - 33.3 mmol/L (10 - 600mg/dL) Meter dimensions (L x W x H) : 82.0 x 45.0 x 15.5 mm (3.22 x 1.77 x 0.61 inch). Display area (L x W) :34.0 x 27.5 mm (1.09 x 1.34 inch), Weight (with battery) : 43.0 ± 5 g, Power supply :1 × CR2032 battery, Memory capacity: 500 test results with date and time, Operating temperature : 6 - 44℃ (43 - 111°F), Operating relative humidity: 10 - 90%, Meter storage conditions :-10 - 60°C (14 - 140°F), Test strip storage conditions :At 4 - 30°C (39 - 86°F), and below 90% relative humidity, Bluetooth version (For GM700SB) :4.0, Accuracy : For glucose concentrations < 5.5 mmol/L (100 mg/dL), 95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements. For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), 95% of measured glucose values within ± 15% of reference measurements. Precision: For glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL) For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5%, Fully compliant with ISO 15197:2013. (For other received credentials, please contact us.)